D-Dimer Microlisa
Test Kit Type: D Dimer Test Kit (ELISA)
Analytical Sensitivity: 25 ng/ml
Total Test Time: 55 Minutes
Shelf life: 15 Month at 2-8°C
Intended Use
D-Dimer Microlisa is designed for in-vitro quantitative determination of D-Dimer in human plasma.
Salient Features
Microwell ELISA Immunoassay for the Quantitative Detection of D-Dimer in Human Plasma
- Based on Sandwich ELISA principle.
- Ready to use Enzyme conjugate.
- Measuring Range: 25 ng/ml to 10000 ng/ml.
- Analytical Sensitivity 25 ng/ml
- Total Test Time :55 minutes.
- Shelf life :15 month at 2-8°C
- Breakaway Microwells provided in 24 test pack.
- Colour coded reagents to monitor procedural steps.
- Convenient pack size: 96 Test.
Principle
D-Dimer Microlisa test is an enzyme immunoassay based on sandwich ELISA. Anti-D-Dimer antibodies has been coated onto the wells of the microwell plate provided. Plasma sample including standards are pipetted into these wells.
During the first incubation, human D-Dimer antigen from samples and coated antibody is react and after washing the enzyme conjugate (anti-D-Dimer antibody HRPO) is added after incubation and washing to remove all the unbounded enzyme conjugate. A substrate solution which acts on the bound enzyme is added to induce a coloured reaction product. The intensity of this coloured product is directly proportional to the concentration of D-Dimer present in the samples.
Ordering Information
Product Name | Pack Size | Catalog No. |
---|---|---|
D-DIMER Microlisa | 96 Tests | IR005096 |
Performance Characteristics (Internal Evaluation) Of D-Dimer Microlisa Test
- Accuracy : The kit has been tested using 25 clinical samples in comparison with commercially available reference immunoassay test kits and the Accuracy co-efficient is 0.995.
- Precision
A) The test results are repeatable in within run precision testing with CV <10%.
B) The test results are reproducible in between run precision testing with CV <15%.